Input Power 115 VAC
Frequency 50/60Hz
Fuses Two 5 x 20 1.6A slow

Dimensions 22 cm (W) x 14 cm (H) x 33 cm (D)
Weight 6.4 kg

Operating temperatures 4 - 30 Degrees Celsius
Environmental conditions 90% Rh, 75-106 kPa

Output range 1 - 600 V (minimum 1 V steps)
1 - 500 mA (minimum 1 mA steps)
Accuracy Voltage: +/- 1% or 2 V
Current: +/- 1% or 2 mA
Output Power 1 to 150 Watt

Process Protection Automatic cross-over
Auto-recovery after power loss
Power overload protection
Current overload protection
"Voltage not reached" protection
Short circuit
Timer Settings 0 - 99 hrs (Minimum step size 1 minute)

Electrical safety Class 1
Installation Category II
EMC Standards EN 55011:1998 and EN 50081-2:1993 and EN 61326:1997, concerning the emission of radiated electromagnetic disturbances

























Home
About Us
Links
Contact Us
EPS 600 (catalogue number 1526)
Return to Previous Page
CreaCon News
Auto Resume After Power Failure
With this new added functionality it is possible to select whether or not to automatically resume the currently running method after a power failure. When auto resume is selected and a power failure occurs the method is automatically resumed and on display a power failure message appears along with the power failure duration time. When auto resume is not selected the T-Cy comes up with a power failure message and asks whether or not resume the method. In the configuration menu the auto resume option can be selected. Also the last occurrence of a power failure is displayed.
Available from September deliveries on, all T-Cy's are provided with software version 1.10 and have a new low noise emission fan build in.